PCR workstations are helps keep contaminants in the air away from biological and genetic samples used in PCR. Workstations are also known as PCR hoods, clean benches, laminar flow hoods, and other names
Technical specification :
1. Internal size (H x W x D) : Minimum 2.5 x 3 x 2 Feet
2. The instrument must be free-standing, Not Tabletop
3. Working bench height: Minimum 2.5 Feet
4. Main body:
a) Cold Rolled steel with anti bacterial coating, corrosion- resistant, durable, and smooth finish.
b) Working Table: 304 stainless steel with satin or brushed finish, corrosion-resistant, hygienic and easy to clean.
5. Front & Side Windows: The front and side windows should be toughened glass with a minimum thickness of 5 mm with UV protection or equivalent safety standards.
6. Laminar system:
a) Pre-filter: Washable Polyester fiber
b) HEPA filter: 99.999% efficiency at 0.3 micron
c) Airflow velocity : 0.3 – 0.5 m/s, speed adjustable
d) Noise : ≤65dB
7. UV Lamp: The UV light should have a power rating of 25-30 W and a wavelength of 254 nm for effective decontamination.
8. LED lamp: 20 W or more
9. Waterproof socket
10. Power supply: 220V±10%, 50/60Hz
11. Warranty: 2 years
Class II, Type B2 cabinets are total exhaust cabinets, widely used in toxicology laboratories and similar applications where chemical effluent is present and clean air is essential